Design and Implementation of Low-Cost Non-Invasive Blood Pressure Measurement Device from Photoplethysmography with Signal Quality AssessmentShow others and affiliations
2026 (English)In: Proceedings - 2026 13th International Conference on Future Internet of Things and Cloud, FiCloud 2026, Institute of Electrical and Electronics Engineers (IEEE) , 2026, p. 341-348Conference paper, Published paper (Refereed)
Abstract [en]
Cardiovascular diseases remain a leading global health concern, necessitating accurate, non-invasive blood pressure monitoring solutions. This study presents a novel photoplethysmography (PPG)-based system that integrates three key innovations: (1) a real-time signal quality assessment (SQA) model employing a one-class SVM classifier to filter unreliable PPG segments, (2) a computationally efficient two-stage neural network (1D U-Net followed by 1D MultiResUNet) for arterial blood pressure (ABP) waveform estimation, and (3) a wearable hardware platform combining a MAX30102 optical sensor and ESP32 microcontroller for portable deployment. The SQA model uses five optimized features to achieve robust motion artifact rejection. The deep learning pipeline reconstructs ABP waveforms with mean absolute errors of 4.7 mmHg (systolic) and 4.3 mmHg (diastolic), complying with the Association for the Advancement of Medical Instrumentation (AAMI) standards. Validated on 40 subjects against reference sphygmomanometer measurements, the system demonstrates 92.5% (systolic) and 95% (diastolic) classification accuracy, with subsecond latency and low-power operation. By addressing critical challenges in motion robustness, computational efficiency, and clinical validation, this work advances the practicality of cuffless BP monitoring for telehealth and resource-limited settings.
Place, publisher, year, edition, pages
Institute of Electrical and Electronics Engineers (IEEE) , 2026. p. 341-348
Series
International Conference on Future Internet of Things and Cloud (FiCloud), ISSN 2996-1017
Keywords [en]
Photoplethysmography, Arterial Blood Pressure, Deep Learning, Signal Quality Assessment, Wearable Devices
National Category
Signal Processing Other Medical Engineering
Research subject
Cyber Security
Identifiers
URN: urn:nbn:se:ltu:diva-119807DOI: 10.1109/FiCloud70576.2026.00054Scopus ID: 2-s2.0-105049591450OAI: oai:DiVA.org:ltu-119807DiVA, id: diva2:2102847
Conference
13th International Conference on Future Internet of Things and Cloud (FiCloud 2026), Granada, Spain, July 20-22, 2026
Note
ISBN for host publication: 979-8-3195-4739-2