Open this publication in new window or tab >>2026 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]
Background: Total hip replacement is one of the most successful orthopaedic interventions; however, wear remains a key factor influencing long-term implant performance and survivorship. Additively manufactured (AM) PEEK materials have emerged as potential candidates for orthopaedic bearing applications, although their tribological and biological performance under clinically representative conditions remains insufficiently understood.
Aim: The aim of this thesis was to evaluate the tribological and biological performance of AM PEEK materials for orthopaedic bearing applications, focusing on the influence of manufacturing, surface finishing, composite reinforcement, wear particle isolation and characterisation. In parallel, the thesis investigated the influence of physiologically representative loading profiles on wear behaviour in conventional metal-on-polymer systems to assess limitations in current preclinical testing approaches.
Methods: This thesis combined a systematic review with experimental evaluation of AM PEEK materials and conventional orthopaedic bearing systems.Physiologically representative loading profiles were applied to conventional metal-on-polymer hip systems using full-scale hip simulator testing to investigate their influence on preclinical wear assessment. The increased wear observed under adverse loading conditions informed subsequent investigations of AM PEEK using multidirectional pin-on-disc articulation testing at elevated contact pressures. These investigations assessed surface topography, wear behaviour and wear particle generation. A vacuum oven stabilisation approach was developed to improve gravimetric assessment of AM PEEK by reducing influence of manufacturing induced porosity on fluid absorption and mass stability. Wear particles were isolated using a phenol-based separation method designed to retain both polyethylene and PEEK particles despite similarities in material and protein density, before characterisation using scanning electron microscopy and in-vitro biological assays.
Key findings: Surface integrity was identified as a critical determinant of tribological performance in AM PEEK, with manufacturing induced porosity and surface condition exerting greater influence than surface roughness alone. A vacuum oven stabilisation approach enabled improved gravimetric assessment of AM PEEK by reducing variability associated with fluid uptake and release.A phenol-based isolation method enabled recovery and characterisation of both polyethylene and PEEK wear particles for biological evaluation. AM PEEK demonstrated biological responses comparable to conventional materials, although composite reinforcement did not improve wear or biological performance.Finally, physiologically representative loading increased wear in conventional metal-on-polymer systems relative to standardised conditions, highlighting the importance of clinically representative preclinical evaluation to reduce the risk of underestimating in-vivo wear.
Interpretation: Collectively, the findings indicate that tribological performance of additively manufactured PEEK is governed more strongly by manufacturing and finish-induced surface integrity than by surface roughness alone. Physiologically representative loading and multidirectional articulation substantially influenced wear behaviour, highlighting limitations of simplified preclinical testing approaches and reinforcing the importance of clinically relevant evaluation frameworks. Furthermore, while AM PEEK demonstrated biological responses comparable to conventional materials, improvements in tribological performance remain important for supporting further biological validation and development of orthopaedic bearing systems.
Conclusions: This thesis demonstrates the importance of clinically relevant tribological evaluation for assessing emerging orthopaedic bearing materials. Physiologically representative loading altered wear behaviour in conventional metal-on-polymer systems, while multidirectional articulation provided greater insight into the tribological performance of AM PEEK materials.While AM PEEK demonstrated biological responses comparable to conventional materials, tribological performance remained strongly dependent on manufacturing and finish-induced surface integrity and was not improved through the introduction of composite reinforcement.Overall, the findings support continued investigation of AM PEEK as a candidate orthopaedic bearing material, while highlighting the need for further optimisation of manufacturing processes and more clinically representative evaluation methodologies prior to translation.
Place, publisher, year, edition, pages
Luleå: Luleå University of Technology, 2026
Series
Doctoral thesis / Luleå University of Technology, ISSN 1402-1544
Keywords
Total hip replacement, Orthopaedic tribology, Polyetheretherketone (PEEK), Additive manufacturing, Wear, Wear particles, Multidirectional articulation, Physiological loading, Surface integrity, Biological response
National Category
Bio Materials Composite Science and Engineering Other Mechanical Engineering Manufacturing, Surface and Joining Technology
Research subject
Machine Elements
Identifiers
urn:nbn:se:ltu:diva-120057 (URN)978-91-8142-167-5 (ISBN)978-91-8142-168-2 (ISBN)
Public defence
2026-12-01, E231, Luleå University of Technology, Luleå, 09:00 (English)
Opponent
Supervisors
2026-10-082026-10-062026-11-13Bibliographically approved